AMX AXT-CV TiltScreen CATP Manual de Instruções Página 138

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 168
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 137
22. CHECKS AND ADJUSTMENTS
130
If the tilt angle shown on the display shifts from tilt angle 0° (zero point), the instrument is not correctly
levelled. This will adversely affect angle measurement.
Perform the following procedure to cancel the tilt zero point error.
PROCEDURE Checking and adjusting
1. Carefully level the OS. If necessary, repeat the
procedures to check and adjust the bubble levels.
2. Select "Inst. cons." in <Configuration>
3. Select "Tilt offset".
4. Level the instrument until the X/Y tilt angles are
±1’. Wait a few seconds for the display to
stabilize, then read the current tilt angle in the X
(sighting) direction and Y (horizontal axis)
direction.
22.2
Tilt Sensor
Vista de página 137
1 2 ... 133 134 135 136 137 138 139 140 141 142 143 ... 167 168

Comentários a estes Manuais

Sem comentários